undefined
+
  • undefined

LAB-EDI-U(10-90LPH)

  • Commodity name: LAB-EDI-U(10-90LPH)
  • Product Description
  • Specification

    Model Number

    Capacity

    Delivery time

    Warranty

    Payment term

    Configuration

    LAB-EDI-U series

    10L/H

    15L/H

    30L/H

    60L/H

    90L/H

    7-10 working days after payment

    1 year

    T/T 100% in advance

    Machine+terminal filter+accessories kit

     

    Product feature 

    ➣ Colored touched screen, humanized operation interface make users completely monitor controller’s working status and real-time data.
    ➣ Built-in filter, with 10 micron 316L stainless steel precision filter screen to remove fine impurities instead of PP cotton filter; manual drain valve, automatic cleaning at any time, soft scraping design for more thorough drainage, transparent shell make filtration effect visual.
    ➣ The long-acting KDF device adopts the original imported filter material . It removes residual chlorine, heavy metals, and inhibits the growth of fungi at the front of water purification. It has a long service life, better protects the post-purification system, and extends lifetime of ultra-purification column 
    ➣ Lifetime of PP cotton, activated carbon filter, RO membrane and ultra purification columns can be set to display the elapsed&rest time of consumables. Automatic reminder for filter replacement to avoid water quality decreasing.
    ➣ Imported UF ultra filtration column ensure meet request for removing pyrogen and RNA.
    ➣ The pipeline adopts a quick-plug interface, the pipe joint passes NSF certification, dual-wavelength UV lamps, effectively sterilizes and reduces TOC, and the original imported 5000D ultra filtration module effectively removes pyrogen.
    ➣ Type U Korean quick-insert filters, used imported filtering materials
    ➣ Inbuilt feed water pressure test gauge to real time monitor tap water pressure and filters blocking.
    ➣ Perfect managing functions of DI &UP water making, circulation system and sterilization.
    ➣ Built-in pressure tank which has inner pouch whose material is anti-bacteria and it is transparent, the inner pouch can be changed anytime.
    ➣ The machine is equipped with pure water transfer pump to ensure the flow rate of water 
    ➣ ABS engineering plastic case, powder coating technology, with water quality and quantity upgrading function, door is open at the side
    ➣ Backlight LCD display,dual waterway online monitoring, which can simultaneously display DI conductivity value,UP resistance value and equipment operating status indication.
    ➣ Menu-type operation, parameters can be set and modified, operating status display and indication light indicate to promptly reflect machine operating status such as high-pressure shutdown, low-pressure alarm, water full shutdown, etc.
    ➣ Alarming for water quality exceeding, automatic filters replacement reminding, water intake time can be set. Take water by touched key, Automatic indication light indicate DI and UP water intake.
    ➣ Equipped with a portable test pen, which can test TDS content, conductance and water temperature anytime, anywhere
    ➣ With RS-232C communication interface output, it can be connected with intelligent instruments with RS-232C interface to realize the transmission of measurement data or print records 
    ➣ Can be equipped with ultra-pure water internal circulation system which keep water quality at the highest level. 
    ➣ Using EDI module, Ensure the purity of water and maintain the life of consumables.
    ➣ 40 L water level tank to store EDI water 

     

    Parameter

    Type classification

    Standard

    Low Endotoxin

    Low TOC

    Comprehensive

    The machine model

    LAB-EDI-U

    LAB-EDI-U-F

    LAB-EDI-U-V

    LAB-EDI-U-VF

    Water rate

    10/15/30/60/90 lph

    The water requirements

    Urban tap water, inlet water pressure: 0.15-0.5mpa, water temperature 5-40℃

    RO water Parameter(ASTM TYPE IV grade water)

    Conductivity

    1-5 us/cm

    resistivity

    0.2-1MΩ.cm@ 25 ℃

    DI water Parameter(ASTM TYPE II grade water)

    Resistivity (25℃)

    10-16 M Ω. cm @ 25 ℃

    Conductivity

    <0.1 μS/cm

    Application

    Preposition unit of ultra pure water system, supply water for biochemical analyzer, electroplating painting, battery charging, chips washing, microbial culture media, chemical&biochemical regent preparation, buffer making, photographic film processing, stability test chamber, meteorological test equipment, hydrogen generator, salt spray test chamber.

    UP water Parameter(ASTM TYPE I grade water)

    Resistivity (25℃)

    18.25 M Ω. Cm @ 25 ℃

    Heavy metal ion

    < 0.01 ppb

    Total organic carbon (TOC) *

    <10 ppb

    <3 ppb

    bacteria

    < 1 cfu/ml *(with terminal filter)

    Particulate matter (> 0.1 microns)

    <1/ml

    Pyrogen (endotoxin)

    N/A

    < 0.001 Eu/ml

    N/A

    < 0.001 Eu/ml

    Ribonuclease (RNases)

    N/A

    <1pg/ml

    N/A

    <1pg/ml

    Deoxyribonucleases

    N/A

    <5pg/ml

    N/A

    <5pg/ml

    Application

    All kinds of physical and chemical testing, biochemical analysis, blood examination, microanalysis, flushing special, HPLC, IC, GC, analysis experiment

    High sensitivity ICP/MS, PCR, PPT level analysis, isotope analysis, molecular biology, life science, CDC, drug control institute, quality inspection institute

    Atomic absorption/emission spectrometry, mass spectrometry, TOC detection, environmental monitoring, biochip development, trace organic matter analysis

    Molecular biology, Life sciences, gene research, cell culture, amino acid analysis, protein purification, toxicology, drug development, medical testing

    Electrical requirements

    100-240V, 50/60Hz, 48-72W

    The system configuration

    Host machine (including 1 set of purification column)+TDS pen + accessory bag

     

    Operation screen display

     

    Filter function

    Front filter

    10 micron 316 stainless steel washable filter, remove minor impurities,extend PP cotton lifetime.

    Granular activated carbon filter

    remove color, odor, residual chlorine and heavy metals

    EDI module

    KDF filter

    Remove residual chlorine,heavy metals, retrain fungi breed, lifetime is very long,   can protect post-positive filters.

    Reverse osmosis membrane

    filter out <1nm pollutants, ions, particles

    PP cotton filter

    remove impurities (greater than 5 micron) such as suspended solid, rust, sludge, colloid, etc.

    Dual wavelength  UV lamp 

    sterilization device, decomposition and photo oxidation OC, lower TOC  (*-V,-VF)

    (Electro-deionzation, referred to as EDI) is a new water treatment method that fills mixed-bed resin between ion exchange membranes and achieves continuous desalination under the action of a DC electric field.

    Ultra-purification column

    ion exchange resin bed, H+ and OH- ion removal

    Ultra filtration membrane

    remove pyrogen,RNase, DNase (*-F,-VF)

    0.2μm terminal filter

    final removal for < 0.2μm bacteria

     

    Process

Keyword: